Proximal landing zone of TEVAR. Are there any limits? A narrative review

Available Online: July, 2025
Page: 67–75

Author for correspondence:

Spyridon N. Mylonas, MD,MSc, PhD, FEBVS
Department of Vascular Surgery, G.H.A. “G. GENNIMATAS”, Mesogeion Av. 154, 11527, Athens, Greece
E-mail: s.mylonas@gna-gennimatas.gr

ISSN 2732-7175 / 2025 Hellenic Society of Vascular and Endovascular Surgery
Published by Rotonda Publications
All rights reserved. https://www.heljves.com

Spyridon N. Mylonas, Vasileios Papavlasopoulos, Vasileios Katsikas, George Geropapas, Alexis Kalamaras, Petros Chatzigakis, George Kopadis
Department of Vascular Surgery, G.H.A. “G. GENNIMATAS”, Athens, Greece

Abstract:

Thoracic endovascular aortic repair (TEVAR) is increasingly utilized for managing thoracic aortic pathologies, necessitating adequate proximal stent graft apposition to healthy aortic wall to optimize outcomes..This review highlights the evolution of TEVAR techniques and devices, emphasizing the importance of anatomical suitability, technical success rates, and the need for a standardized approach to LSA reperfusion.Ultimately, careful patient selection, device choice, and interdisciplinary collaboration are critical to enhancing procedural outcomes and reducing complications in TEVAR.

Thoracic endovascular aortic repair (TEVAR) has gained widespread use in the treatment of acute and chronic pathologies of the thoracic aorta. Optimal outcomes in TEVAR are dependent upon obtaining adequate proximal apposition of stent graft against a healthy aortic wall.1, 2 According to the instructions for use of the available endografts, this should be at least 15-20 millimeters long. (Table 1) In the proximal thoracic aorta, this may require coverage of one or more of the branches of the aortic arch, most commonly the left subclavian artery (LSA) (26-40%).3-5 However, coverage of the LSA during endovascular repair of the thoracic aorta has been identified in the literature as a significant modifiable risk factor for: cerebrovascular ischemia, stroke (anterior circulation), vertebral ischemia, ischemia of the upper extremities.3, 4, 6-8 There are also some special cases that make reperfusion of the LSA imperative, such as:

-Presence of a patent coronary bypass between the left internal mammary artery and coronary arteries.

-Termination of the left vertebral artery on the posterior inferior cerebellar artery.

-Absent, hypoplastic or occluded right vertebral artery.

-Presence of a patent arteriovenous dialysis shunt in the left upper limb.

-Planned extensive coverage (20 cm) of the descending thoracic aorta.

-Previous infrarenal aortic surgery with concomitant ligation of lumbar and middle sacral arteries.

-Occlusion of the internal iliac artery.5, 9, 10

Based on the above, all current guidelines of the European Society of Vascular Surgery, the Society of Vascular Surgery and the American Heart Association recommend prior or simultaneous reperfusion of the LSA, when coverage is necessary for adequate proximal landing zone achievement, in order to prevent neurological ischemic complications.9, 11, 12 Traditionally, reperfusion of the LSA was performed with a carotid-subclavian bypass. (Figure 1) Alternatively, LSA is transposed to the left common carotid artery (LCCA).13 (Figure 2) Although this technique eliminates the need for a synthetic graft and does not require secondary central embolization, it is contraindicated in cases of early take-off of the left vertebral artery and in the presence of a patent coronary bypass from the left internal mammary artery. Adopting all these open LSA reperfusion techniques, an excellent patency rate of >97% has been described at five years follow-up.14, 15

Therefore, arises the question whether there is a need for alternative methods of reperfusion of the LSA. Indeed, in obese patients, open reperfusion of the LSA becomes technically demanding and is associated with an increased risk of local complications, the incidence of which is not negligible; i.e bleeding, (10-20%), peripheral nerve injury (9-25%) and lymph leakage, although relatively rare (3-5%), may require more sophisticated treatment techniques.15-17 (Figure 3)

This was realized by the team of Lawrence-Brown in Liverpool, who first described in 2004 the in-situ creation of a fenestration in a thoracic dacron endograft using the hard tip of a guide wire and the sequential use of cutting balloons so that a covered stent could be placed and blood flow to the LSA could be maintained.18 While in 2009, the team of Frank Arco from Texas described for the first time the use of an ultraviolet laser catheter to create a fenestration in a Talent thoracic endograft and, after balloon dilation, placed an iCast covered stent in a patient with traumatic thoracic aortic injury.19 A recent meta-analysis on the use of laser to create fenestrations during the endovascular treatment of aortic arch pathologies identified 6 studies with a total of 247 patients, almost 81% of whom underwent surgery on an emergency or urgent setting. The pooled technical success rate was estimated at 98%, the pooled 30-day mortality rate was estimated at 3.2% and the pooled estimate for stroke was estimated at 4.5%. Regarding the mid-term results, a target-vessel patency rate of 100% and a reintervention rate of 6% during a median follow-up period of 12 months was recorded.20 The recently published early results of the multicenter LIFE (Laser In Situ Fesnestrated Endograft) confirmed the feasibility of the technique with a perioperative mortality and neurologic risk profile that is comparable to standard techniques, but with a modest (9.8% within 15 months) reintervention rate.21

In 2021, the Fu-through™ in situ needle fenestration system in combination with the Fustar™ steerable guided sheath and the Ankura™ thoracic endograft (Lifetech Scientific, Shenzhen, China) were introduced in Europe and received CE mark. This system allows stabilizing and centralizing the 20G puncture needle, which is designed with 3 puncture depths and is convenient for 0.018” guidewire to go through and PTA balloon can be used for rapid dilation. In a single-center retrospective study on 52 TEVAR patients, an anatomic feasibility rate of 61.5% (32/52) was found based on the aortic arch morphology. The LSA angulation was identified to be the most important and limiting anatomical constraint.22 The experience so far, mainly from China, has shown excellent results in terms of technical success, 30-day mortality and the occurrence of cerebrovascular events.23-27 (Table 2)

Various other techniques for in-situ fenestration of thoracic endografts such as the use of recanalization catheters, or wire connected to electrodiathermy have been suggested. However, all of them are characterized by a lack of standardization, a combination of many materials and technologies and of course the absence of long-term results. Thus, the need for a standardized endovascular method for reperfusion of the LSA arises.

The solution to this need aspires to be the single-branch stent-grafts. Currently available in Europe is the GORE TBE™ (W. L. Gore & Associates, Newark, DE, USA) (MDR CE mark 2024), which is based on the well-known TAG platform and has a retrograde internal branch for the subclavian artery of 8 mm at a distance of 20 to 25 mm from the central end, a 12 mm branch at a distance of 40 mm from the central end. It includes a bridging endograft with diameters of 8 or 12 mm and a length of 6 cm and an aortic extension of 36 to 46 mm. (Figure 5) So far, this endograft has been used mainly in the United States of America, where it has been available in clinical studies since 2014 with excellent results.28-35 (Table 3) However, it should be noted that, as with all endografts, there is a limitation of anatomical suitability. In a retrospective single center study of 210 patients who underwent TEVAR in zone 2, only 38.1% of patients met the anatomical criteria for a central landing zone. The left carotid artery – LSA distance being the most frequent factor of anatomic suitability violation.36

The second branched thoracic endograft currently available in Europe -in a custom-made platform- is the Castor™ (Lombard Medical Limited, in partnership with MicroPort Endovastec, Newark UK). It is a PTFE-coated nitinol scaffold endograft with a single branch 6-14 mm in diameter and 25 or 30 mm in length at a distance of 5 to 30 mm from the central tip. (Figure 6) The endograft was initially available in China with the indication of type B aortic dissection. A recent meta-analysis identified 11 studies with a total of 415 patients.37 Technical success was recorded at 97.5%, no postoperative strokes were observed, while the 30-day mortality was almost 1%, while the patency of the LSA branch at 12 months was found to be 95%. Recently the first experience from Europe was published with encouraging results.38 The second generation of Castor™, Cratos™, which has a redesigned delivery mechanism with a 2 Fr smaller sheath, was recently put into clinical trials in Europe and Japan (ongoing CREATION clinical trial).

Alternatives are also currently available as custom-made stent-grafts: the fenestrated endograft Cook Arch Fen™ (by COOK, Bloomington, IN, USA), the Najuta™ (by B-Kawasumi Laboratories, Inc., Kanagawa, Japan) and the Relay™ (by Terumo Aortic, Inchinnan, UK). All 3 endografts have shown high technical success rates with a stroke risk of 3.9% to 7.5%.39-41 However, they set as an anatomical prerequisite the presence of a 2 cm sealing zone (with diameter <40mm) in the mid-aortic arch, which is not always possible, especially in cases of type III aortic arch. In such cases, it is necessary to shift the proximal sealing zone centrally, towards the ascending aorta. Thus, branched aortic arch endografts, the Relay™ (by Terumo Aortic, Inchinnan, UK), Cook Arch Branch™ (by COOK, Bloomington, IN, USA) and the NEXUS™ (by Endospan, Herzlia, Israel) with central sealing zone in the ascending aorta, were developed. (Figure 7)

The multicenter postmarket study for Nexus, which included 28 patients, showed excellent technical success rates (100%), and extremely low stroke rates (3.6%), but in 29% of patients a reintervention was necessary at 36 months.42, 43 Similarly, a multicenter study on the Terumo Relay branched endograft, which included 43 patients showed a 30-day mortality of 9%, and 7% disabling stroke rate.44

Regarding the COOK endograft, the version with the 2 branches, the initial experience was rather disappointing with a technical success of only 84% and a 30-day mortality at 13% and cerebrovascular accidents at approximately 16%.45 These results were attributed to the required learning curve and, indeed, in the following series excellent technical success rates were observed, as well as acceptable 30-day mortality rates. However, the risk of stroke remained significant.45-47 Recently, the international experience with the version with the 3 branches was also published. An improvement was found in terms of the risk of stroke, but the need for secondary interventions was estimated at 31% during a median follow-up period of 3.2 months.48 Interstingly, a recent study from Hamburg found that the proximal landing zone in the native ascending aorta is associated with high stroke rates (13.5%, 7.9% major) and high need for reinterventions (24-month freedom from rentervention 46.4%).49

It should, however, be mentionted that the anatomical suitability of these endografts ranges from 10% to 50%, depending on the population under examination. (Table 5) All of studies identified the short length (<5cm) and the large diameter (>40mm) of the ascending aorta as the main factors for limiting the anatomical suitablity.50-54 Indeed, the presence of a mechanical aortic valve, the short length of the ascending aorta and its severe angulation constitute significant challenges for the use of branched aortic arch endografts. Tsilimbaris et al. proposed the use of a small tip, 35 mm, and its passage lateral to the metallic leaflet, in order to preserve the function of the other one in cases of patients with a metallic aortic valve.55 A recently published multicentre experience on these patients showed excellent technical success rates. However, it should be mentioned that, the included patients were highly selected and the participating centers were experienced high volume aortic centres.56 Moreover, in patients with short ascending aorta length, endografts with 3 inverted branches can be used.57 It should be noted, however, that these endografts are not readily available to everyone, so in selected cases the chimney technique, the in-situ fenestration technique or surgeon-modified endografts can be applied.

Recently, have been published efforts to expand the proximal landing zone in the aortic root. These are based in the combination of fenestrated or branched endografts along with percutaneous aortic valves with encouraging early outcomes for highly selected patients. Long-term follow-up data are, however, needed to assess the persistence of the seal and ongoing durability of this novel technique. 58-62

In summary, the proximal sealing zone represents the most important factor related to early and long-term results after TEVAR and ensuring its adequate length is the main challenge. Several different technologies have expanded the boundaries of what can be achieved through endovascular techniques worldwide. While the long-term results of the use of branched endografts are pending, most technical limitations are being eliminated and a satisfactory level of technical success has been noted with F/BTEVAR. However, this comes at the cost of increased stroke and reoperation rates. Therefore, patient selection, respecting the anatomical requirements, judicious choice of the most appropriate device considering the pros and cons of each system, in a high-volume centre and by an interdisciplinary team are the keys to minimize complications and achieve an adequate proximal landing zone.

References

  1. Boufi M, Aouini F, Guivier-Curien C, Dona B, Loundou AD, Deplano V, et al. Examination of factors in type I endoleak development after thoracic endovascular repair. J Vasc Surg. 2015;61(2):317-23.
  2. Yoon WJ, Mell MW. Outcome comparison of thoracic endovascular aortic repair performed outside versus inside proximal landing zone length recommendation. J Vasc Surg. 2020;72(6):1883-90.
  3. Luehr M, Etz CD, Berezowski M, Nozdrzykowski M, Jerkku T, Peterss S, et al. Outcomes After Thoracic Endovascular Aortic Repair With Overstenting of the Left Subclavian Artery. Ann Thorac Surg. 2019;107(5):1372-9.
  4. Waterford SD, Chou D, Bombien R, Uzun I, Shah A, Khoynezhad A. Left Subclavian Arterial Coverage and Stroke During Thoracic Aortic Endografting: A Systematic Review. Ann Thorac Surg. 2016;101(1):381-9.
  5. Matsumura JS, Rizvi AZ, Society for Vascular S. Left subclavian artery revascularization: Society for Vascular Surgery Practice Guidelines. J Vasc Surg. 2010;52(4 Suppl):65S-70S.
  6. Chen X, Wang J, Premaratne S, Zhao J, Zhang WW. Meta-analysis of the outcomes of revascularization after intentional coverage of the left subclavian artery for thoracic endovascular aortic repair. J Vasc Surg. 2019;70(4):1330-40.
  7. Batubara EAD, Nugraha RA, Amshar M, Siddiq T, Indriani S, Adiarto S. Ischemic Complications Following Thoracic Endovascular Aortic Repair with and Without Revascularization of Left Subclavian Artery: A Systematic Review and Meta-Analysis. Annals of Vascular Surgery. 2022;86:417-27.
  8. Rizvi AZ, Murad MH, Fairman RM, Erwin PJ, Montori VM. The effect of left subclavian artery coverage on morbidity and mortality in patients undergoing endovascular thoracic aortic interventions: a systematic review and meta-analysis. J Vasc Surg. 2009;50(5):1159-69.
  9. Isselbacher EM, Preventza O, Hamilton Black J, 3rd, Augoustides JG, Beck AW, Bolen MA, et al. 2022 ACC/AHA Guideline for the Diagnosis and Management of Aortic Disease: A Report of the American Heart Association/American College of Cardiology Joint Committee on Clinical Practice Guidelines. Circulation. 2022;146(24):e334-e482.
  10. Alanezi T, Altoijry A, AlSheikh S, Al-Mubarak H, Alhamzah M, Alomran F, et al. Predicting the need for subclavian artery revascularization in thoracic endovascular aortic repair: A systematic review and meta-analysis. J Vasc Surg. 2024;80(3):922-36 e5.
  11. Czerny M, Schmidli J, Adler S, van den Berg JC, Bertoglio L, Carrel T, et al. Editor’s Choice – Current Options and Recommendations for the Treatment of Thoracic Aortic Pathologies Involving the Aortic Arch: An Expert Consensus Document of the European Association for Cardio-Thoracic Surgery (EACTS) & the European Society for Vascular Surgery (ESVS). Eur J Vasc Endovasc Surg. 2019;57(2):165-98.
  12. Upchurch GR, Jr., Escobar GA, Azizzadeh A, Beck AW, Conrad MF, Matsumura JS, et al. Society for Vascular Surgery clinical practice guidelines of thoracic endovascular aortic repair for descending thoracic aortic aneurysms. J Vasc Surg. 2021;73(1S):55S-83S.
  13. Bianco V, Sultan I, Kilic A, Aranda-Michel E, Cuddy RJ, Srivastava A, et al. Concomitant left subclavian artery revascularization with carotid-subclavian transposition during zone 2 thoracic endovascular aortic repair. The Journal of Thoracic and Cardiovascular Surgery. 2020;159(4):1222-7.
  14. Protack CD, Smith A, Moennich LA, Hardy D, Lyden SP, Farivar BS. Midterm outcomes of subclavian artery revascularization in the setting of thoracic endovascular aortic repair. J Vasc Surg. 2020;72(4):1222-8.
  15. Voigt SL, Bishawi M, Ranney D, Yerokun B, McCann RL, Hughes GC. Outcomes of carotid-subclavian bypass performed in the setting of thoracic endovascular aortic repair. J Vasc Surg. 2019;69(3):701-9.
  16. Konstantinou N, Debus ES, Vermeulen CFW, Wipper S, Diener H, Larena-Avellaneda A, et al. Cervical Debranching in the Endovascular Era: A Single Centre Experience. Eur J Vasc Endovasc Surg. 2019;58(1):34-40.
  17. D’Oria M, Kärkkäinen JM, Tenorio ER, Oderich GS, Mendes BC, Shuja F, et al. Perioperative Outcomes of Carotid-Subclavian Bypass or Transposition versus Endovascular Techniques for Left Subclavian Artery Revascularization during Nontraumatic Zone 2 Thoracic Endovascular Aortic Repair in the Vascular Quality Initiative. Ann Vasc Surg. 2020;69:17-26.
  18. McWilliams RG, Murphy M, Hartley D, Lawrence-Brown MM, Harris PL. In situ stent-graft fenestration to preserve the left subclavian artery. J Endovasc Ther. 2004;11(2):170-4.
  19. Murphy EH, Dimaio JM, Dean W, Jessen ME, Arko FR. Endovascular repair of acute traumatic thoracic aortic transection with laser-assisted in-situ fenestration of a stent-graft covering the left subclavian artery. J Endovasc Ther. 2009;16(4):457-63.
  20. Houerou TL, Nana P, Pernot M, Guihaire J, Gaudin A, Lerisson E, et al. Systematic Review on In Situ Laser Fenestrated Repair for the Endovascular Management of Aortic Arch Pathologies. J Clin Med. 2023;12(7).
  21. Ullery BW, Alie-Cusson F, Magee GA, Han SM, Stern JR, Barleben A, et al. Laser In Situ Fenestrated Endograft (LIFE) Repair of Complex Aortic Arch Pathology: Early Outcomes from the Multicenter LIFE Registry. Journal of Vascular Surgery.
  22. Piffaretti G, Franchin M, Gnesutta A, Gatta T, Piacentino F, Rivolta N, et al. Anatomic Feasibility of In-Situ Fenestration for Isolate Left Subclavian Artery Preservation during Thoracic Endovascular Aortic Repair Using an Adjustable Needle Puncturing System. J Clin Med. 2023;13(1).
  23. Fan B, Fang K, Tian C, Fang J, Chen D, Zhao J, et al. In Situ Fenestration and Carotid-Subclavian Bypass for Left Subclavian Artery Revascularization During Thoracic Endovascular Aortic Repair. Cardiovasc Intervent Radiol. 2024;47(6):717-27.
  24. Li G, Li M, Dong Z, Gu J, Liu H, Xu X, et al. In situ needle fenestration for aortic arch conditions during thoracic endovascular aortic repair. BMC Cardiovasc Disord. 2024;24(1):625.
  25. Usai MV, Austermann M. Experience with the Ankura Thoracic Stent Graft and In-situ Fenestration for the Left Subclavian Artery with the Fu-Through Needle – a Technical Overview and Comparison to Similar Endovascular Techniques. Zentralbl Chir. 2023;148(5):425-8.
  26. Shu X, Xu H, Wang E, Wang L, Guo D, Chen B, et al. Midterm Outcomes of an Adjustable Puncture Device for In Situ Fenestration During Thoracic Endovascular Aortic Repair. Eur J Vasc Endovasc Surg. 2022;63(1):43-51.
  27. Yu Z, Hu S, Wang D, Yang T, Lang D. Early and midterm outcomes of in situ fenestration via adjustable puncture needle for Ankura aortic stent graft: A single-center experience. Vascular. 2024;32(5):964-72.
  28. Dake MD, Brinkman WT, Han SM, Matsumura JS, Sweet MP, Patel HJ, et al. Outcomes of endovascular repair of aortic aneurysms with the GORE thoracic branch endoprosthesis for left subclavian artery preservation. J Vasc Surg. 2022;76(5):1141-9 e3.
  29. Dake MD, Fischbein MP, Bavaria JE, Desai ND, Oderich G, Singh MJ, et al. Evaluation of the Gore TAG thoracic branch endoprosthesis in the treatment of proximal descending thoracic aortic aneurysms. J Vasc Surg. 2021;74(5):1483-90 e2.
  30. Patel HJ, Dake MD, Bavaria JE, Singh MJ, Filinger M, Fischbein MP, et al. Branched Endovascular Therapy of the Distal Aortic Arch: Preliminary Results of the Feasibility Multicenter Trial of the Gore Thoracic Branch Endoprosthesis. Ann Thorac Surg. 2016;102(4):1190-8.
  31. Squiers JJ, DiMaio JM, Schaffer JM, Baxter RD, Gable CE, Shinn KV, et al. Surgical debranching versus branched endografting in zone 2 thoracic endovascular aortic repair. J Vasc Surg. 2022;75(6):1829-36 e3.
  32. Desai ND, Wang GJ, Brinkman W, Coselli J, Taylor B, Patel H, et al. Outcomes of a Novel Single-Branched Aortic Stent Graft for Treatment of Type B Aortic Dissection. Ann Thorac Surg. 2025;119(4):826-34.
  33. DiLosa K, Harding J, Donde N, Silverman D, Heafner T. Symptomatic carotid stenosis in Eagle syndrome managed with traditional endarterectomy and styloidectomy. J Vasc Surg Cases Innov Tech. 2024;10(1):101372.
  34. DiLosa KL, Manesh M, Kanamori LR, Chan M, Magee GA, Fleischman F, et al. Multi-center experience with an off-the-shelf single retrograde thoracic branch endoprosthesis for acute aortic pathology. J Vasc Surg. 2025;81(4):839-46.
  35. Liang NL, Dake MD, Fischbein MP, Bavaria JE, Desai ND, Oderich GS, et al. Midterm Outcomes of Endovascular Repair of Aortic Arch Aneurysms with the Gore Thoracic Branch Endoprosthesis. Eur J Vasc Endovasc Surg. 2022;64(6):639-45.
  36. Blitti C, Die-Loucou J, Jazayeri A, Bouchot O, Bamde CC, Steinmetz E. Anatomic Suitability of Left Subclavian Artery Off-The-Shelf Branch Graft in a Real-World Setting. Ann Vasc Surg. 2025;122:792-800.
  37. Yao S, Chen X, Liao Y, Ding G, Li D, Qin G, et al. Systematic review and meta-analysis of type B aortic dissection involving the left subclavian artery with a Castor stent graft. Front Cardiovasc Med. 2022;9:1052094.
  38. Zolnierczuk M, Rynio P, Rybicka A, Glowinski J, Milnerowicz A, Pormanczuk K, et al. Initial Multicenter Experience With the New Castor Arch Branched Device in Europe: A Middle-term Results Study. J Endovasc Ther. 2024:15266028241304307.
  39. Tsilimparis N, Prendes CF, Rouhani G, Adam D, Dias N, Stana J, et al. Pre-Loaded Fenestrated Thoracic Endografts for Distal Aortic Arch Pathologies: Multicentre Retrospective Analysis of Short and Mid Term Outcomes. Eur J Vasc Endovasc Surg. 2021;62(6):887-95.
  40. Sica S, Pratesi G, Rossi G, Ferraresi M, Lovato L, Volpe P, et al. Proximal sealing in the aortic arch for inner curve disease using the custom Relay scalloped and fenestrated stent graft. J Vasc Surg. 2024;80(5):1317-25 e2.
  41. Isernia G, Simonte G, Orrico M, Silingardi R, Gaggiano A, Covic T, et al. Preliminary results from an Italian National Registry on the outcomes of the Najuta fenestrated aortic arch endograft. J Vasc Surg. 2023;77(5):1330-8 e2.
  42. Planer D, Elbaz-Greener G, Mangialardi N, Lindsay T, D’Onofrio A, Schelzig H, et al. NEXUS Arch: A Multicenter Study Evaluating the Initial Experience With a Novel Aortic Arch Stent Graft System. Ann Surg. 2023;277(2):e460-e6.
  43. D’Onofrio A, Lachat M, Mangialardi N, Antonello M, Schelzig H, Chaykovska L, et al. Three-year follow-up of aortic arch endovascular stent grafting with the Nexus device: results from a prospective multicentre study. Eur J Cardiothorac Surg. 2022;63(1).
  44. Czerny M, Berger T, Kondov S, Siepe M, Saint Lebes B, Mokrane F, et al. Results of endovascular aortic arch repair using the Relay Branch system. Eur J Cardiothorac Surg. 2021;60(3):662-8.
  45. Haulon S, Greenberg RK, Spear R, Eagleton M, Abraham C, Lioupis C, et al. Global experience with an inner branched arch endograft. J Thorac Cardiovasc Surg. 2014;148(4):1709-16.
  46. Spear R, Haulon S, Ohki T, Tsilimparis N, Kanaoka Y, Milne CP, et al. Editor’s Choice – Subsequent Results for Arch Aneurysm Repair with Inner Branched Endografts. Eur J Vasc Endovasc Surg. 2016;51(3):380-5.
  47. Tsilimparis N, Detter C, Law Y, Rohlffs F, Heidemann F, Brickwedel J, et al. Single-center experience with an inner branched arch endograft. J Vasc Surg. 2019;69(4):977-85 e1.
  48. Tenorio ER, Oderich GS, Kolbel T, Dias NV, Sonesson B, Karelis A, et al. Multicenter global early feasibility study to evaluate total endovascular arch repair using three-vessel inner branch stent-grafts for aneurysms and dissections. J Vasc Surg. 2021;74(4):1055-65 e4.
  49. Nana P, Spanos K, Panuccio G, Rohlffs F, Detter C, von Kodolitsch Y, et al. Branched and fenestrated endovascular aortic arch repair in patients with native proximal aortic landing zone. J Vasc Surg. 2024;80(3):621-9 e3.
  50. Burke CR, Kratzberg JA, Yoder AD, Steele NZ, Aldea GS, Sweet MP. Applicability of the Zenith Inner Branched Arch Endograft. J Endovasc Ther. 2020;27(2):252-7.
  51. Smorenburg SPM, Montesano M, Hoogteijling TJ, Truijers M, Symersky P, Jansen EK, et al. Anatomic Suitability for Branched Thoracic Endovascular Repair in Patients with Aortic Arch Pathological Features. J Am Heart Assoc. 2020;9(20):e016695.
  52. Benfor B, Prendes CF, Peterss S, Stavroulakis K, Stana J, Pichlmaier M, et al. Anatomical analysis of the aortic arch and feasibility study of double and triple branched thoracic endografts in the treatment of isolated arch pathologies. Eur J Cardiothorac Surg. 2022;62(3).
  53. Exelmans W, Mufty H, Maleux G, Verbrugghe P, Fourneau I. Anatomical feasibility of an endovascular aortic arch repair with the NEXUS endograft in patients treated with a frozen elephant trunk procedure for aortic arch pathology. CVIR Endovasc. 2023;6(1):10.
  54. Leone N, Bartolotti LAM, Baresi GF, Silingardi R, Resch TA, Gennai S. Anatomical suitability for branched endovascular aortic arch repair and balloon-expandable bridging stent grafts in a cohort of patients previously treated with a hybrid approach. J Vasc Surg. 2024;79(2):198-206 e15.
  55. Konstantinou N, Peterss S, Stana J, Rantner B, Banafsche R, Pichlmaier M, et al. Passing a Mechanical Aortic Valve With a Short Tip Dilator to Facilitate Aortic Arch Endovascular Branched Repair. Journal of Endovascular Therapy. 2021;28(3):388-92.
  56. Konstantinou N, Jakimowicz T, Haulon S, Pichlmaier M, Abisi S, Pedro LM, et al. Outcomes after Endovascular Arch Repair in Patients with a Mechanical Aortic Valve: Results from a Multicentre Study. Eur J Vasc Endovasc Surg. 2025;69(2):263-70.
  57. Nana P, Houérou TL, Guihaire J, Gaudin A, Fabre D, Haulon S. Early Outcomes on Triple-Branch Arch Device With Retrograde Left Common Carotid Branch: A Case Series. J Endovasc Ther. 2025;32(3):802-12.
  58. Ghoreishi M, Chahal D, Shah A, Kang J, Hirsch J, Tran D, et al. First-in-Human Endovascular Aortic Root Repair (Endo-Bentall) for Acute Type A Dissection. Circ Cardiovasc Interv. 2023;16(10):e013348.
  59. Felipe Gaia D, Bernal O, Castilho E, Baeta Neves Duarte Ferreira C, Dvir D, Simonato M, et al. First-in-Human Endo-Bentall Procedure for Simultaneous Treatment of the Ascending Aorta and Aortic Valve. JACC Case Rep. 2020;2(3):480-5.
  60. Leshnower BG, Duwayri YM, Nicholson WJ, Ueyama H, Gleason PT, Shekiladze N, et al. Endo-Bentall Procedure Using Off-the-Shelf Catheter Devices to Repair an Aorto-Atrial Fistula. Circ Cardiovasc Interv. 2023;16(4):e012848.
  61. Ghoreishi M, Shah A, Chahal D, Kang J, Gupta A, Taylor BS, et al. Endo-Bentall repair: Early results and feasibility of a physician-constructed Endo-Bentall device. J Thorac Cardiovasc Surg. 2025;170(3):639-49.
  62. Shehab M, Prendes CF, Wanhainen A, Lindblom R, James S, Mani K. The Endo-Bentall Procedure in the Treatment of Ascending Aorta Pathologies: A Scoping Review. Eur J Vasc Endovasc Surg. 2025.

    Privacy Preferences

    When you visit our website, it may store information through your browser from specific services, usually in the form of cookies. Here you can change your Privacy preferences. It is worth noting that blocking some types of cookies may impact your experience on our website and the services we are able to offer.

    Click to enable/disable Google Analytics tracking code.
    Click to enable/disable Google Fonts.
    Click to enable/disable Google Maps.
    Click to enable/disable video embeds.
    Our website uses cookies, mainly from 3rd party services. Define your Privacy Preferences and/or agree to our use of cookies.